Display range
Digital display (3 digits) 10 kΩ to 15 TΩ
Analogue display 100 kΩ to 1 TΩ
Short circuit/charge current
5 mA @ 5 kV
Capacitor charge time
<1.5 seconds per μF at 5mA to 5 kV
Capacitor discharge time
<120ms per μF to discharge from 5000 V to 50 V
Capacitance measurement (above 500 V)
10 nF to 50 μF (dependent on measurement voltage)
Capacitance measurement accuracy (23°C)
±5% ±5 nF
Voltage output accuracy (0°C to 30°C)
+4%, -0%, ±10 V of nominal test voltage at 1 GΩ load
Current measurement range
0.01 nA to 5 mA
Current measurement accuracy (23°C)
±5% ±0.2 nA at all voltages
Interference rejection
S1-552 1 mA per 250 V up to a maximum of 2 mA
Timer range
Up to 99 minutes and 59 seconds from start of test
15 second minimum setting for test voltage ≥1000 V
30 second minimum setting for test voltage <1000 V
Memory capacity
32kB
Test regimes
Auto IR, PI, DAR, SV and DD.
Interface
RS232 and USB
Data store
Voltage, test time, leakage current, resistance, PI, DAR, DD,
capacitance and time constant
Real time output
Serial, once per second of test voltage, current and resistance
Operating temperature
-10°C to 50°C
Storage temperature
-25°C to 65°C
Ingress protection (lid closed)
IP65
Humidity
90% RH non-condensing at 40°C
Safety
Meets the requirements of EN61010-1:2001 CATIII 300 V
EMC
Meets the requirements of EN61326-1:1998 for use in heavy
industrial areas.